Descripción
|
|
---|---|
The deviation of calibration coefficients from five cup anemometer models over time was analyzed. The analysis was based on a series of laboratory calibrations between January 2001 and August 2010. The analysis was performed on two different groups of anemometers: (1) anemometers not used for any industrial purpose (that is, just stored); and (2) anemometers used in different industrial applications (mainly in the field?or outside?applications like wind farms). Results indicate a loss of performance of the studied anemometers over time. In the case of the unused anemometers the degradation shows a clear pattern. In the case of the anemometers used in the field, the data analyzed also suggest a loss of performance, yet the degradation does not show a clear trend. A recalibration schedule is proposed based on the observed performances variations. | |
Internacional
|
Si |
JCR del ISI
|
Si |
Título de la revista
|
Energies |
ISSN
|
1996-1073 |
Factor de impacto JCR
|
1,865 |
Información de impacto
|
|
Volumen
|
5 |
DOI
|
10.3390/EN5051664 |
Número de revista
|
5 |
Desde la página
|
1664 |
Hasta la página
|
1685 |
Mes
|
MAYO |
Ranking
|