Initial performance evaluation of a high resolution Albira small animal positron emission tomography scanner with monolithic crystals and depth-of-interaction encoding from a user¿s perspective
Año:2009

Áreas de investigación
  • Industria electrónica

Datos
Descripción
The user evaluation of the Albira small animal positron emission tomography scanner is presented. The system has a ¿80 mm × 40 mm field of view. In the center of the field of view, it has 2.49% sensitivity, with 33% solid angle coverage, and 0.88% sensitivity for a mouse phantom. The maximum of noise equivalent counts is 14.6 kcps at 6.0 MBq for a mouse phantom. The scanner employs an innovative crystal design of eight 40(50) × 40(50) × 9.8 mm3 LYSO tapered monolithic crystals forming detector modules read by position-sensitive photomultipliers. This design allows for easy depth-of-interaction readout. The system saturates at 6.7 MBq for the mouse phantom. As an example of the application of the Albira system, some preliminary results from a study of ischemic stroke on animal models with 18-FDG co-registered with MRI images are shown.
Internacional
Si
JCR del ISI
Si
Título de la revista
MEASUREMENT SCIENCE AND TECHNOLOGY
ISSN
0957-0233
Factor de impacto JCR
1,493
Información de impacto
Volumen
20
DOI
10.1088/0957-0233/20/10/104011
Número de revista
0
Desde la página
104011
Hasta la página
6p
Mes
SEPTIEMBRE
Ranking

Esta actividad pertenece a memorias de investigación

Participantes

Grupos de investigación, Departamentos, Centros e Institutos de I+D+i relacionados
  • Creador: Grupo de Investigación: Tecnología de imágenes biomédicas
  • Centro o Instituto I+D+i: Centro de tecnología Biomédica CTB
  • Departamento: Ingeniería Electrónica