Memorias de investigación
Artículos en revistas:
Low-resource synchronous coincidence processor for positron emission tomography
Año:2011

Áreas de investigación
  • Equipo médico,
  • Tecnología electrónica y de las comunicaciones,
  • Industria electrónica

Datos
Descripción
We developed a new FPGA-based method for coincidence detection in positron emission tomography. Themethod requires low device resources and no specific peripherals in order to resolve coincident digital pulses within a time window of a few nanoseconds. This method has been validated with a low-end Xilinx Spartan-3E and provided coincidence resolutions lower than 6 ns. This resolution depends directly on the signal propagation properties of the target device and the maximum available clock frequency, therefore it is expected to improve considerably on higher-end FPGAs
Internacional
Si
JCR del ISI
Si
Título de la revista
Nucl. Instrum. Methods Phys. Res. Sect. A-Accel. Spectrom. Dect. Assoc. Equip
ISSN
0168-9002
Factor de impacto JCR
1,207
Información de impacto
Volumen
648
DOI
10.1016/j.nima.2010.11.170
Número de revista
Desde la página
S199
Hasta la página
S201
Mes
AGOSTO
Ranking

Esta actividad pertenece a memorias de investigación

Participantes

Grupos de investigación, Departamentos, Centros e Institutos de I+D+i relacionados
  • Creador: Grupo de Investigación: Tecnología de imágenes biomédicas
  • Centro o Instituto I+D+i: Centro de tecnología Biomédica CTB
  • Departamento: Ingeniería Electrónica