Memorias de investigación
Ponencias en congresos:
Characterization and Test of a Data Acquisition System for PET
Año:2011

Áreas de investigación
  • Equipo médico,
  • Tecnología electrónica y de las comunicaciones,
  • Industria electrónica

Datos
Descripción
A small Positron Emission Tomography demonstrator based on LYSO slabs and Silicon Photomultiplier matrices is under construction at the University and INFN of Pisa. In this paper we present the characterization results of the read-out electronics and of the detection system. Two SiPM matrices, composed by 8 x 8 SiPM pixels, 1.5 mm pitch, have been coupled one to one to a LYSO crystals array. Custom Front-End ASICs were used to read the 64 channels of each matrix. Data from each Front-End were multiplexed and sent to a DAQ board for the digital conversion; a motherboard collects the data and communicates with a host computer through a USB port. Specific tests were carried out on the system in order to assess its performance. Futhermore we have measured some of the most important parameters of the system for PET application.
Internacional
Si
Nombre congreso
IEEE Nuclear Science Symposium Conference
Tipo de participación
960
Lugar del congreso
Valencia, España
Revisores
Si
ISBN o ISSN
1082-3654
DOI
10.1109/NSSMIC.2011.6154124
Fecha inicio congreso
23/10/2011
Fecha fin congreso
29/10/2011
Desde la página
621
Hasta la página
625
Título de las actas
IEEE Nuclear Science Symposium Conference Record

Esta actividad pertenece a memorias de investigación

Participantes

Grupos de investigación, Departamentos, Centros e Institutos de I+D+i relacionados
  • Creador: Grupo de Investigación: Tecnología de imágenes biomédicas
  • Centro o Instituto I+D+i: Centro de tecnología Biomédica CTB
  • Departamento: Ingeniería Electrónica